Skip to Main Content »

Search Site

Gemini Multi-Channel Controller

Double click on above image to view full picture

Zoom Out
Zoom In

More Views

  • Gemini Multi-Channel Controller

MODEL GEMINI · Dual-Channel · with digital timer on one channel

Email to a Friend

Availability: In stock.

$1,425.00

Quick Overview

Available for purchase with or without system accessories.
Just the controller $1425.00
Complete system $1625.00


Gemini controls TWO independent reactions! Dual-channel controller features two temperature controllers in a single cabinet to regulate two independent reactions. The Gemini is among J-KEM’s most versatile controllers featuring high power outlets and a 100 hour digital timer (on channel 1) to turn heating OFF (or ON) at a user entered time. Includes over-temperature protection circuits with audible alarms, dual temperature displays, selectable operating modes, double fused, and 100% solid state design. Contains J-KEM’s patented power control computer which provides 0.1o C regulation of anything.



  • 0.1º C Regulation & Display

  • NIST traceable

  • Advanced PID algorithm


2 Year warranty


Also available as 230VAC, CE approved Model GEMINI-CE


Academic and volume discounts apply


Complete system includes: one controller, two Teflon thermocouples (12" long, 1/8" diameter), two connecting cords, and two 14/20 joint adapters.

Product Description

New in 2014


USB port with free KEM-Net software and KEM-IO feature


At the heart of J-KEM’s 200-Series controllers is a new, high speed microprocessor that performs 3 functions:


1. Power Regulation J-KEM’s original power control computer is replaced by a next generation microprocessor. The power control computer is J-KEM’s patented technology that adjusts power to the heater 2048 times per second resulting in 0.1o C regulation.


2. USB Communications PC communications and free KEM-Net software enable remote PC control, GLP/GMP compliant data collection, and multi-temperature ramps built in an Excel-like table.


3. KEM-IO Allows the controller to turn other pieces of equipment On/Off, change the temperature set point, or Start/Stop heating based on external inputs, temperature, or time.


 


Examples:


Turn heating off when the hood door is opened. Turn it back on when closed.


Open a 120 Vac cooling valve if the reaction exotherms to 110°C.


When the temperature reaches 85°C, start a reagent addition pump or stirrer.

* Required Fields